Fixing distorted images in 3D chemical mapping

New methods help correct errors in three-dimensional time-of-flight secondary ion mass spectrometry images.

Three-dimensional time-of-flight secondary ion mass spectrometry (3D ToF-SIMS) builds chemical volumes. It does this by stacking images acquired between sputter cycles.

Standard reconstruction assumes a flat starting surface and a single sputter rate. This causes distortions in interface positions and depth scales for samples with topography or different sputter rates.

Researchers tested three correction methods using a Pt/TaOx/Ta/Pt/Ti crossbar memristor device. The first method uses 3D ToF-SIMS data alone. It restores relative topography but the depth scale remains unconstrained below the last calibrated interface.

The second method uses atomic force microscopy (AFM) topography before and after the profile. This fixes the start and end surfaces but assumes uniform sputtering in between. This fails near the device where surroundings change material.

The third method uses AFM topography at intermediate stages. This follows material changes and reproduces the device structure. It requires more instrument time and repeated imaging of a partially sputtered surface.

A decision flowchart and a checklist help readers choose, validate, and report a correction.

Key facts

  • 3D ToF-SIMS builds chemical volumes by stacking images from sputter cycles.
  • Data-only correction requires no additional instrumentation.
  • Multi-stage AFM correction reproduces the structure of the device.
  • Standard reconstruction can yield distorted interface positions and depth scales.